Determination of antimony by using tungsten trap atomic absorption spectrometry
Date
2008Author
Titretir, SerapKenduezler, Erdal
Arslan, Yasin
Kula, Ibrahim
Bakirdere, Sezgin
Ataman, O. Yavuz
Metadata
Show full item recordAbstract
An electrically heated tungsten coil was used as a trap in the determination of antimony. The technique consists of three steps. Initially, SbH(3) is formed by hydride generation procedure; then the analyte species in vapor form are transported to W-coil trap heated at 370 degrees C. Following the preconcentration step, the trap is heated to 895 degrees C; analyte species are revolatilized and transported to the flame-heated quartz atom cell where atornization and the formation of signal take place. The experimental parameters were optimized both for trap and no-trap studies. The most important experimental parameters are concentrations of HCI and NaBH4 solutions. H2 and At gas flow rates. and collection and revolatilization temperatures of W-coil. Accuracy was tested using a certified reference material, waste water EU-L-1. Limit of detection for the system is 16 ng 1- 1 using a sample of 36 ml collected in 4.0 min. Enhancement factor in sensitivity was 17. (c) 2008 Elsevier B.V. All rights reserved.