dc.identifier.citationMetin, M., Aksoy, K., Pekmezci, A., Deniz, A., 2023. Comparison of the blood gas analyser Edan i15 Vet with the blood gas analyser Gem Premier 3000 for the analysis of blood gases, electrolytes, hemoglobin and hematocrit in lactating Holstein. Medycyna Weterynaryjna 79, 6728–2023.. doi:10.21521/mw.6728en_US
dc.identifier.issn0025-8628
dc.identifier.urihttp://dx.doi.org/10.21521/mw.6728
dc.identifier.urihttps://hdl.handle.net/20.500.12809/10498
dc.description.abstractA point-of-care blood gas analyser, Edan i15 Vet (EDAN), was compared with a benchtop blood gas analyser, GEM Premier 3000 (GEM). GEM and EDAN were used to analyse whole blood from 123 lactating Holsteins within one month of calving for blood gases, electrolytes, hematocrit and hemoglobin. EDAN and GEM showed significant linear correlations for blood gases, electrolytes, Hct and Hb. The 95% confidence intervals (CI) of intercept in Passing-Bablok regressions included zero in pCO2, bicarbonate, TCO2, BE ecf, BE B, hematocrit, and pH, but not for Na+, K+, pO2 and sO2. The CI of the slope included 1.0 for Na+, K+, pCO2, bicarbonate, TCO2, BE ecf, BE B, hematocrit, and pH, but not for pO2, sO2, and hemoglobin. The Bland-Altman plots between EDAN and GEM showed a bias of 1.4% for Na+, 2.4% for K+, -1.6% for pCO2, 3.0% for pO2, -5.3% for bicarbonate, 2.8% for SO2, -7.3% for TCO2, 10.4% for Hct, 21.2% for Hb, -25.1% for BE B and -38.5% for BE ecf. The biases in the analysis of certain estimated parameters were much higher (> 5%) than for measured parameters except for Hct. Parity did not correlate with blood gas parameters but blood pH correlated negatively with K+, pCO2 and positively with pO2, TCO2, sO2, bicarbonate, BE ecf and BE B. The postpartum time correlated positively with pCO2, TCO2, BE and bicarbonate and negatively with Hct and Hb values. Reference values (2.5-97.5% quartiles) were determined for each parameter. Conclusively, EDAN can be used interchangeably with GEM for the analysis of blood pH, K+, pCO2 and HCO3- as they show acceptable moderate agreement. However, they did not agree for other parameters such as TCO2, BE ecf, BE B, Hb, Hct, sO2, Na+, and pO2, therefore, reference values for each parameter were set for GEM and EDAN.en_US
dc.item-language.isoengen_US
dc.publisherPOLISH SOC VETERINARY SCIENCES EDITORIAL OFFICEen_US
dc.relation.isversionof10.21521/mw.6728en_US
dc.item-rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectBlood gasesen_US
dc.subjectEdanen_US
dc.subjectGemen_US
dc.subjectHolsteinen_US
dc.titleComparison of the blood gas analyser Edan i15 Vet with the blood gas analyser Gem Premier 3000 for the analysis of blood gases, electrolytes, hemoglobin and hematocrit in lactating Holsteinsen_US
dc.item-typearticleen_US
dc.contributor.departmentMÜ, Milas Veteriner Fakültesi, Temel Bilimler Bölümüen_US
dc.contributor.authorID0000-0002-1402-4594en_US
dc.contributor.authorID0000-0003-0149-6688en_US
dc.contributor.authorID0000-0003-4020-0069en_US
dc.contributor.institutionauthorMert, Metin
dc.contributor.institutionauthorAksoy, Kemal
dc.contributor.institutionauthorPekmezci, Aytaç
dc.identifier.volume79en_US
dc.identifier.issue2en_US
dc.identifier.startpage83en_US
dc.identifier.endpage90en_US
dc.relation.journalMEDYCYNA WETERYNARYJNA-VETERINARY MEDICINE-SCIENCE AND PRACTICEen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US


Bu öğenin dosyaları:

Thumbnail

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster