dc.description.abstractPurpose: The purpose of this study was to compare the central corneal thickness (CCT) measurements of four noncontact devices in healthy eyes. Materials and Methods: In a sample of 45 healthy controls, CCT was measured using an optical biometer (IOLMaster 700) based on swept-source optical coherence tomography (SS-OCT), high-resolution rotating Scheimpflug camera system (Pentacam HR), spectral-domain OCT (SD-OCT) device with an anterior segment module (Spectralis), and noncontact pachymetry (NCP) device (Topcon TRK-2P). Agreement among the devices was analyzed using mean differences (i. e., bias) and Bland-Altman analysis with 95% limits of agreement (LoA). Results: Mean CCT measurements were 537.5 +/- 47.5 mu m for SS-OCT optical biometer, 532.3 +/- 43.8 mu m for Scheimpflug system, 521.3 +/- 44.7 mu m for SD-OCT device, and 518.0 +/- 43.1 mu m for NCP (P < 0.001). The SD-OCT device and NCP showed the closest agreement, with a bias of 2.6 mu m (95% LoA, -3.6-8.8 mu m), whereas the SS-OCT optical biometer and NCP showed the least agreement, with a bias of 18.7 mu m (95% LoA, -2.1-39.5 mu m). Bias was 16.1 mu m (95% LoA, -3.1-35.3 mu m) for SS-OCT optical biometer and SD-OCT, 5.1 mu m (95% LoA, -6.8-17.0 mu m) for SS-OCT optical biometer and Scheimpflug system, 10.9 mu m (95% LoA, -15.1-36.9 mu m) for SD-OCT device and Scheimpflug system, and 13.6 mu m (95% LoA, -5-32.2 mu m) for Scheimpflug system and NCP. Conclusions: SS-OCT optical biometer overestimates CCT measurements compared to Scheimpflug system, SD-OCT device, and NCP. Given mean differences and range variations in CCT measurements between devices, SS-OCT optical biometer and Scheimpflug system are interchangeable as are SD-OCT and NCP.en_US
dc.item-language.isoengen_US
dc.publisherMedknow Publications & Media Pvt Ltden_US
dc.item-rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectAnterior Segment Optical Coherence Tomographyen_US
dc.subjectCentral Corneal Thicknessen_US
dc.subjectNoncontact Pachymetryen_US
dc.subjectScheimpflug Analysis Systemen_US
dc.subjectSwept-Source Optical Biometeren_US
dc.titleComparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technologyen_US
dc.item-typearticleen_US
dc.contributor.departmentMÜ, Eğitim ve Araştırma Hastanesien_US
dc.contributor.institutionauthorÖzyol, Erhan
dc.contributor.institutionauthorÖzyol, Pelin
dc.identifier.doi10.4103/ijo.IJO_618_16
dc.identifier.volume65en_US
dc.identifier.issue6en_US
dc.identifier.startpage461en_US
dc.identifier.endpage465en_US
dc.relation.journalIndian Journal of Ophthalmologyen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US


Bu öğenin dosyaları:

Thumbnail

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster