dc.contributor.author | Şimşek, Cem | |
dc.contributor.author | Kaya, Cansu | |
dc.contributor.author | Karalezli, Aylin | |
dc.date.accessioned | 2022-11-07T11:20:32Z | |
dc.date.available | 2022-11-07T11:20:32Z | |
dc.date.issued | 2022 | en_US |
dc.identifier.citation | Şimşek C, Kaya C, Karalezli A. Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry. Turk J Ophthalmol. 2022 Oct 28;52(5):318-323. doi: 10.4274/tjo.galenos.2021.22465. PMID: 36317802; PMCID: PMC9631500. | en_US |
dc.identifier.issn | 2147-2661 / 2149-8709 | |
dc.identifier.uri | https://hdl.handle.net/20.500.12809/10368 | |
dc.description.abstract | Objectives: To assess the central corneal thickness (CCT) with 5 different devices, evaluate the repeatability of the devices, and determine the possible relationship between thickness values and sex.
Materials and methods: The study included 308 eyes of 154 patients (76 women, 78 men) between the ages of 18-30 who presented to the Ophthalmology Clinic of Muğla Sıtkı Koçman University Training and Research Hospital. Autorefractor (Topcon, Japan), ultrasound pachymetry (UP) (Ceniscan, USA), high-resolution Pentacam (Oculus, USA), anterior segment-optical coherence tomography (AS-OCT) (Optovue, USA), and Spectralis AS-OCT (Heidelberg, Germany) measurements were assessed.
Results: The mean age of the study participants was 23.2±0.2 years and the mean CCT was 540±14.1 μm, with no statistically significant difference in CCT between sexes (p>0.05). Mean CCT values were 557.0±26.7 μm with the autorefractor, 543.6±32.9 μm with UP, 533.8±30.2 μm with the Oculus Pentacam, 519.8±30.1 μm with Optovue AS-OCT, and 547.5±31.6 μm with Heidelberg AS-OCT. Pairwise comparisons between devices showed that the Optovue AS-OCT gave significantly lower CCT measurements than the autorefractor and Heidelberg AS-OCT device (p=0.027 and p=0.033, respectively). The coefficient of repeatability for autorefractor, UP, high-resolution Pentacam, Optovue AS-OCT, and Heidelberg AS-OCT CCT measurements were 1.51%, 2.46%, 3.72%, 2.57%, and 3.34%, respectively.
Conclusion: Measurements made with five different devices showed that CCT was comparable and clinically usable. However, it was determined that the Optovue AS-OCT showed lower CCT values compare to other devices. When compared in terms of repeatability, it was found to be lower in the Pentacam than other devices. | en_US |
dc.item-language.iso | eng | en_US |
dc.publisher | Galenos Publishing House | en_US |
dc.relation.isversionof | 10.4274/tjo.galenos.2021.22465 | en_US |
dc.item-rights | info:eu-repo/semantics/openAccess | en_US |
dc.subject | Central corneal thickness | en_US |
dc.subject | Cornea | en_US |
dc.subject | Pachymetry | en_US |
dc.subject | Optical coherence tomography | en_US |
dc.title | Comparison of Central Corneal Thickness Measurements with Four Different New Devices and Ultrasound Pachymetry | en_US |
dc.item-type | article | en_US |
dc.contributor.department | MÜ, Tıp Fakültesi, Cerrahi Tıp Bilimleri Bölümü | en_US |
dc.contributor.authorID | 0000-0003-1316-4656 | en_US |
dc.contributor.institutionauthor | Şimşek, Cem | |
dc.contributor.institutionauthor | Kaya, Cansu | |
dc.contributor.institutionauthor | Karalezli, Aylin | |
dc.identifier.volume | 52 | en_US |
dc.identifier.issue | 5 | en_US |
dc.identifier.startpage | 318 | en_US |
dc.identifier.endpage | 323 | en_US |
dc.relation.journal | Turkish Journal of Ophthalmology | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |